0X001047
2D SAPE INVALID CAL
Check the Diameter potentiometer connector and cable.
If the problem still persists:
Check the Diameter potentiometer
If the problem still persists:
Check the Integrated Display.
If the problem still persists:
Call Snap On
C Codes C80, C81, C92, C115, C84, C88
0X001050 SIDE CAL BAR
Perform the calibration C115 with weight in the correct side.
If the problem still persists:
Call Snap On
C Codes C64, C115, C84, C88
0X001051 SIDE CAL USER
Perform the User calibration with weight in the correct side.
If the problem still persists:
Call Snap On
C Codes
0X001052 NO CAL USER
Perform the User calibration
If the problem still persists:
Call Snap On
C Codes
0X001053 FAIL CAL USER
Perform the User calibration
If the problem still persists:
Call Snap On
C Codes
0X001054 VIRTUAL DIMENSION OUT OF RANGE
Perform the calibration
If the problem still persists:
Check the transducers wiring.
If the problem still persists:
Check the transducers.
If the problem still persists:
Check the IBP box.
If the problem still persists
Call Snap On
C Codes C66, C86, C115, C84, C88