52
0X001055 CAL1 100G WEIGHT LOW
Perform the calibration with C115.
If the problem still persists:
Check the transducers wiring
If the problem still persists:
Check the transducers adjustment.
If the problem still persists:
Check the IBP box.
If the problem still persists:
Call Snap On
C Codes C66, C75, C103, C104, C115, C84, C88.
0X001056 CAL2 100G WEIGHT LOW
Perform the calibration with C115.
If the problem still persists:
Check the transducers wiring
If the problem still persists:
Check the transducers adjustment.
If the problem still persists:
Check the IBP box.
If the problem still persists:
Call Snap On
C Codes C66, C75, C103, C104, C115, C84, C88.
0X001057 CAL1 0G WEIGHT HIGH
Perform the calibration with C115.
If the problem still persists:
Check the transducers wiring
If the problem still persists:
Check the transducers adjustment.
If the problem still persists:
Check the IBP box.
If the problem still persists:
Call Snap On
C Codes C66, C75, C103, C104, C115, C84, C88.
0X001058 CAL2 0G WEIGHT HIGH
Perform the calibration with C115.
If the problem still persists:
Check the transducers wiring
If the problem still persists:
Check the transducers adjustment.
If the problem still persists:
Check the IBP box.
If the problem still persists:
Call Snap On
C Codes C66, C75, C103, C104, C115, C84, C88.