Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes
36 / 56
NOTE
If a malfunction occurs in the amalgamate separator during "Weekly" clean-
ing, the workflow is discontinued.
Run the "Weekly" cleaning programme:
At initial start-up
Before, during and after prolonged downtimes (more than 4 weeks)
Over the weekend
During vacations
After treatment of patients at high risk
Tipp: Place the tumbler on the tumbler holder first to prevent the water from
splashing.
4 Tap "Weekly" to start the cleaning programme.
ð At the start of the active phase, the display of the dentist element shows
the following message and a beep is issued.
4 During the exposure phase, do not place hoses in holders and, if applicable,
switch off the treatment centre before weekends and extended downtimes.
4 The "Continue" button is shown once the exposure phase is completed.
Press "Continue" to close the hygiene programme. Alternatively, the treat-
ment centre can be switched off at this time.