Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes
37 / 56
4 Wait for the cleaning programme to be completed.
4 Or switch the treatment centre off.
ð A progress bar indicates the cleaning steps.
ð The following message is displayed after completion of the cleaning pro-
gramme:
4 After completion of the "Weekly" cleaning programme, place hoses in hold-
ers.
If the treatment centre was switched off during the exposure phase, the pro-
gramme is continued automatically once the treatment centre is switched on
again.
Intensive germ reduction during prolonged down-times
(more than 4 weeks)
During extended down-times, e.g. semester breaks at university, "Weekly"
cleaning must be run during the down-time. The germ reduction liquid remains
inside the system during the down-time.