EasyManua.ls Logo

KaVo uniQa - Automatic Cleaning Programmes: Weekly Programme Details; Programme Start Confirmation Steps; Weekly Cleaning Programme: Higher OXYGENAL 6 Concentration; OXYGENAL 6 Requirements for Weekly Programme

KaVo uniQa
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Care instructions KaVo uniQa
2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes
41 / 56
Cleaning programme, "Weekly"
The "Weekly" cleaning programme runs an intensive germ reduction at a higher
OXYGENAL 6 concentration.
The following amount of OXYGENAL 6 is required:
Size of water bottle Capacity OXYGENAL 6
1.5 l 65 ml
Using the OXYGENAL 6 Dosing device, 13 dosages at setting 1.5 l are required.
Also refer to:
3.3 Replenishing OXYGENAL 6 when using the water bottle,Page52
The following message is displayed during the cleaning programme:
Preparation of water bottle with higher OXYGENAL 6 concentration
Confirm the completion of the preparation on the dentist element.
4 Press the "Yes" button to confirm that the water bottle containing the re-
quired concentration was attached.