26
Intensity Control
·
·
·
·
Intensity is related to the time it takes to reach the target therapy level after the initiation of
therapy.
The lower the intensity setting, the slower the target therapy level will be reached.
It is recommended that new patients begin therapy at the lowest intensity setting as this
allows for slower increase of negative pressure once the foam is compressed in the
wound.
The intensity can remain at the minimum setting throughout the entire length of treatment, if
desired.
IntensityIntensity
Intensity
Low
Medium High
Press to change levels. Green crescent changes with each setting.
Default setting is: Low
From the Clinician Mode Home screen, press ‘ ’, then to access
th
Therapy then ‘ ’, ‘ ’
e Intensity button.
Next Settings
Settings changed manually take immediate effect when therapy is on.
Settings (cont.)
Pressure Settings
Exit
Pressure
125 mmHg
+
_
Press ‘ ’ to return to the ‘Settings’ screen.Exit
From the Clinician Mode Home screen, press ‘ ’, then
to access th
Therapy then ‘ ’, ‘ ’, then
e Pressure screen.
Next Settings
‘Pressure’
Settings changed manually take immediate effect when therapy is on.
Use the buttons to change to the desired pressure.
Pressure can be set from 25 to 200 mmHg in increments of 25 mmHg.
+
_
D setting is: 125 mmHgefault