Battery Low Alert
To resolve this alert:
1. Connect therapy unit to wall outlet using ActiV.A.C. Power Supply to recharge battery. An
light next to the bottom-left of the Touch Screen User Interface will indicate charging is
underway. Refer to the Battery Charging section of this manual for more information.
2. Press on this screen to return to the ‘Home Screen’.
®
Amber
‘Exit’
This alert screen appears approximately two (2) hours before
the battery power level is too low to support continued operation
of the ActiV.A.C. Therapy Unit.
This alert will be accompanied by a single audible tone.
®
Battery Low; charging required
Battery Low Alert
Battery Low
Continuous
0 mmHg
Exit
This alarm screen appears approximately minutes
before the battery power level is too low to support continued
operation of the
thirty (30)
ActiV.A.C. Therapy Unit.
This alarm will be accompanied by a repeating audible tone.
This alarm may be silenced for two (2) minutes during
troubleshooting by pressing .
®
‘Audio Pause’
Battery Critical Alarm
On/Off
On/Off
If not, press to restart therapy.
To resolve this alarm:
1. Connect therapy unit to wall outlet using ActiV.A.C. Power Supply to recharge battery. An
light next to the bottom-left of the Touch Screen User Interface will indicate charging is
underway. Refer to the Battery Charging section of this manual for more information.
2. Press ‘ ’ on this screen to return to the ‘Home Screen’.
®
Amber
Reset
3. Ensure therapy is on by confirming that the green crescent is lit on the button.
Audio Pause
Reset
Battery Critical
Battery critically low, charge immediately
Battery Low
V.A.C. Therapy continues; however, if this alarm is not resolved within thirty (30) minutes,
therapy will be interrupted.
®
WARNING: Under clinician supervision, replace V.A.C. Dressing with alternate
dressing if therapy is interrupted for more than two hours.
®
V.A.C. Therapy continues.
®
Alerts and Alarms (cont.)
41
®