This alarm screen appears when the ActiV.A.C. Therapy Unit
detects that the canister is full and should be replaced.
This alarm will be accompanied by a repeating audible tone.
This alarm may be silenced for two (2) minutes during
troubleshooting by pressing .
®
‘Audio Pause’
Alerts and Alarms (cont.)
Canister Full Therapy Interrupted Alarm
Reset
Canister Full
Therapy Interrupted
WARNING: If Canister is full, replace and press ‘Reset’.
If not full, press ‘Cancel’. Press ‘?’ for more information
Cancel
Audio Pause
WARNING: Under clinician supervision, replace V.A.C. Dressing with alternate
dressing if therapy is interrupted for more than two hours.
®
Graduated Marks
A full canister is approximately 300 ml.