EasyManua.ls Logo

KCI ActiV.A.C. - Specifications

KCI ActiV.A.C.
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications*
Dimensions:
Weight (with empty canister attached): ~1.08 kg
Pressure Options: 25 to 200 mmHg
(3.3 to 26.6 kPa)
Therapy Delivery Modes: Continuous or Intermittent
Canister Volume: ~300 ml
Battery Run Life: Approximately 14 hours,
depending on settings.
Battery Charge Time: Approximately 6 hours from
a fully discharged state.
External Power Supply Output: 12V, 3.3 A
Patient & Enclosure Leakage Current: <100 Microamps
Storage Conditions
Temperature Range: (-4°F) to 140°F)
Relative Humidity Range: 0-95% non-condensing
Operating Conditions
Temperature Range: 41°F) to (104°F)
5
IEC Classification
Medical Equipment
Equipment not suitable for use in presence of flammable anesthetic mixture with
air, oxygen or nitrous oxide.
Type B, Applied Part
Class II
IPX0
*Specifications subject to change without notice.
19.3 x 15.2 x 6.4 cm
(7.6” W x 6” H x 2.5” D)
(~2.4 lbs)
External Power Supply Input: 100-240VAC .72A @ 115VAC
47-63 Hz
-20°C 60°C (
5°C ( 40°C
Relative Humidity Range: 0-9 % non-condensing
Altitude Range: 0 to 4267 m (0 to 14,000 feet)
Optimum Performance: 0 to 2438 m (0 to 8,000 feet)
Electrical:
Environmental Conditions:
57
®
A
c
t
I

Table of Contents

Other manuals for KCI ActiV.A.C.

Related product manuals