Calibration 14-15
9. Note and record the DMM voltage reading, then calculate the current from that reading
and 4Ω resistor value. Send the following command using that calculated current as the
parameter:
:CAL:PROT:STEP14 <Calculated
_Current>
For example, with a current value of 1.894A, the command would appear as follows:
:CAL:PROT:STEP14 1.894
10. Note and record a new DMM voltage reading, then again calculate the current from the
v
oltage and resistance. Send the calculated current value as the parameter for the
following command:
:CAL:PROT:STEP15 <Calculated
_Current>
11. Connect the characterized 3kΩ
resistor in place of the 4Ω resistor (Figure 14-3).
12. Make sure the DMM DC v
olts function and auto-range are still selected.
13. Send the following command to output approximately 5mA for 5mA full-scale
calibration:
:CAL:PROT:STEP16
14. Note and record the DMM voltage reading, then calculate the
current from the voltage
reading and actual 3kΩ resistance value. Send that current value as the parameter for the
following command:
15. :CAL:PROT:STEP17 <Calcu
lated_Current>
Step 4: Program calibration date
Use following command to set the calibration date:
:CAL:PROT:DATE <
yyyy>, <mm>, <dd>
Note that the year, month, and date must be separated by commas. The allowable range for
the year
is from 1998 to 2097, the month is from 1 to 12, and the date is from 1 to 31.
Step 5: Save calibration constants and lock out calibration
Calibration is now complete, so you can store the calibration constants in EEROM by sending
the following command:
:CAL:PROT:SAVE
NOTE Calibration will be temporary unless you send the SAVE command. Calibration data
will
not be saved if (1) calibration is locked, (2) invalid data exists, or (3) all steps
were not completed in the proper sequence. In that case, the unit will revert to
previous calibration constants.
After saving constants, lock out calibration by sending:
:CAL:PROT:LOCK
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2302

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2302 and is the answer not in the manual?

Keithley 2302 Specifications

General IconGeneral
BrandKeithley
Model2302
CategoryPower Supply
LanguageEnglish

Related product manuals