3-26 Pulse Current Measurements
Figure 3-7
125
425
625
825
975
725
525
325
155
0
100
300
400
500
600
700
900
0.1
0.2 0.3 0.4 0.5
TIME (seconds)
300
TLEV1
TLEV2
TLEV3
TLEV4
TLEV5
TLEV9
TLEV8
TLEV7
TLEV6
900
TLEV mA
(rising edge)
mA TLEV
(falling edge)
125
425
625
825
975
725
525
325
155
100
300
400
500
600
700
900
300
TLEV1
TLEV2
TLEV3
TLEV4
TLEV5
TLEV9
TLEV8
TLEV7
TLEV6
900
TLEV mA
(rising edge)
mA TLEV
(falling edge)
0
2.1
2.2 2.3 2.4 2.5
0.6
0.7
1.8 1.9 2.0
Sample :STEP Pulse measurement
Pulse sequences — rising and falling
Consider the pulse form in Figure 3-8. This pulse form has three falling (DOWN) level steps
followed by three rising (UP) level steps. Since these steps rise and fall to the same steady state
current, acti
ve steps need to be designated as 6 UP and 0 DOWN to measure the step level cur-
rent. If DOWN steps are specified then, the step level current measured will be the steady state
current.
Figure 3-8
Pulse form with rise and fall steps
3 Down
3 Up
350mA
600mA
875mA
725mA
425mA
600μsec
275mA
4msec
Steady state current