Front Panel Operation
4-65
4.12.2 Switching mainframe A settings (7001)
Scan List = 1!1-1!10, 2!1-2!10, M1,
NOTE
M1 is a “dummy” channel configured to open all channels. This lets the last
channel in a mainframe open before closing the first channel in the other
mainframe.
Card Delay = 0.0
Pole Mode = 2-Pole
Scan Configuration:
Arm Control: Spacing TrigLink
Input Trigger Line 5
Output Trigger Line 6
Count 1*
Control Source
Scan Control: Spacing TrigLink
Input Trigger Line 3
Output Trigger Line 4
Count 2
Control Source
Channel Control: Spacing TrigLink
Trigger Link Mode Asynchronous*
Input Trigger Line 1*
Output Trigger Line 2*
Count ScanList*
Control Source*
*RESET default setting