EasyManuals Logo
Home>Keithley>Switch>7001

Keithley 7001 User Manual

Keithley 7001
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #171 background imageLoading...
Page #171 background image
IEEE-488 Reference
5-52
5.10.10
*RST
reset
To reset the Model 7001.
*RST
When the *RST command is sent, the Model 7001 performs the following operations:
1. Returns the Model 7001 to the *RST default conditions (see Table 5-5).
2. Cancels all pending commands.
3. Cancels response to any previously received *OPC and *OPC? commands.
The following instrument states are not affected by this command:
IEEE-488 address
Standard Event Status Register
Status Byte Register
OUTPUT 707; “*RST” ! Return 7001 to *RST defaults
Purpose
Format
Description
Programming
example
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
BrandKeithley
Model7001
CategorySwitch
LanguageEnglish

Related product manuals