IEEE-488 Reference
5-52
5.10.10
*RST
reset
To reset the Model 7001.
*RST
When the *RST command is sent, the Model 7001 performs the following operations:
1. Returns the Model 7001 to the *RST default conditions (see Table 5-5).
2. Cancels all pending commands.
3. Cancels response to any previously received *OPC and *OPC? commands.
The following instrument states are not affected by this command:
IEEE-488 address
Standard Event Status Register
Status Byte Register
OUTPUT 707; “*RST” ! Return 7001 to *RST defaults
Purpose
Format
Description
Programming
example