IEEE-488 Reference
5-60
5.10.15 *TRG trigger
To issue a GPIB trigger to the Model 7001.
*TRG
The *TRG command is used to issue a GPIB trigger to the Model 7001. It has the same effect
as a group execute trigger (GET). The *TRG command is an overlapped command, while GET
is not. See *OPC, *OPC? and *WAI for more information.
The *TRG command is used as an arm, scan and/or channel event to control the scan cycle. The
Model 7001 will react to this trigger if GPIB is the programmed control source. The control
source is programmed from the SCPI :TRIGger Subsystem (see paragraph 5.18).
OUTPUT 707; “*TRG” ! Trigger 7001
Purpose
Format
Description
Programming
example