IEEE-488 Reference
5-83
[:ROUTe] subsystem
Power-up Saved power-on setup
*RST Off
:SYSTem:PRESet Off
:SCHannel? Query status of Single Channel
Short-form formats: :conf:sch?
Response message: 0 (off) or 1 (on)
The Single Channel mode (when enabled) is used to prevent more than one channel from clos-
ing at a time. When a single channel is closed, the previous closed channel will open. This en-
sures that even with random channel closures, only one channel can be closed at a time.
Toggling the Single Channel mode from off to on causes all channels to open. For complete de-
tails on Single Channel, see Section 4.
10 OUTPUT 707; “:conf:sch on; sch?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Enables Single Channel and then queries its status.
Line 20 Addresses the 7001 to talk.
Line 30 Displays the status of Single Channel (1; on).
:CPAir <b>
[:ROUTe]:CONFigure:CPAir <b> Control Card Pair
<b> = 0 or OFF Disable Card Pair.
= 1 or ON Enable Card Pair.
:conf:cpa <b>
Power-up Saved power-on setup
*RST Disables Card Pair
:SYSTem:PRESet Disables Card Pair
:CPAir? Query status of Card Pair
Short-form format: :conf:cpa?
Response message: 0 (off) or 1 (on)
Defaults
Query
command
Description
Programming
example
5.13.7