IEEE-488 Reference
5-131
5.18.8
Trigger subsystem
:SIGNal
:ARM[:SEQuence[1]][:LAYer[1]]:SIGNal Bypass arm control source
:ARM[:SEQuence[1]]:LAYer2:SIGNal Bypass scan control source
:TRIGger[:SEQuence[1]]:SIGNal Bypass channel control source
:arm:sign
These action commands are used to bypass the specified control source and are used when you
do not wish to wait for the programmed event. Keep in mind that the instrument must be waiting
for the appropriate event when the command is sent. Otherwise, an error will occur and this
command will be ignored.
When this command is sent, the specified control source will be bypassed. This will allow op-
eration to continue. Only the layer’s source event between is bypassed; any :DELay time will
be observed.
Note that :SIGNal is not an instrument setup command. As soon as this command is executed,
it will attempt to bypass the control source.
OUTPUT 707; “:trig:sign” ! Bypass channel control source.
TCONfigure commands
:PROTocol <name>
:TRIGger[:SEQuence[1]]:TCONfigure:PROTocol <name> Specify Trigger Link protocol
<name> = ASYNchronous Asynchronous Trigger Link mode
= SSYNchronous Semi-synchronous Trigger Link mode
:trig:tcon:prot <name>
Power-up Saved power-on setup
*RST ASYNchronous
:SYSTem:PRESet ASYNchronous
:PROTocol? Query programmed Trigger Link protocol
Short-form format: :trig:tcon:prot?
Response message: ASYN or SSYN
Short-form
format
Description
Programming
example
5.18.9
Parameters
Short-form
format
Defaults
Query
commands
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com