F
SCPI Conformance Information
F-1
Introduction
The Model 7001 complies with SCPI version 1991.0.
Table F-1
Syntax of SCPI confirmed commands implemented by Model 7001
Command Description
:DISPlay
[ :WINDow[1]]
:TEXT
:DATA <a>
:STATe <b>
:STATe?
:WINDow2
:TEXT
:DATA <a>
:STATe <b>
:STATe?
Path to locate message to top portion of display
Path to control user text messages:
DeÞne ASCII message ÒaÓ using up to 20 characters.
Enable (1 or ON) or disable (0 or OFF) message.
Request message status (enabled or disabled).
Path to locate message to bottom portion of display:
Path to control user text messages:
DeÞne ASCII message ÒaÓ using up to 32 characters.
Enable (1 or ON) or disable (0 or OFF) message.
Request message status (enabled or disabled).
[:ROUTe]
:CLOSe <list>
:STATe?
:OPEN <list>|ALL
:SCAN <list>
:POINts?
Path and command to close speciÞed channels:
Request channels that are closed.
Open speciÞed (or all) channels.
Path and command to specify channel list:
Request number of channels in the scan list.