EasyManuals Logo
Home>Keithley>Switch>7001

Keithley 7001 User Manual

Keithley 7001
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #342 background imageLoading...
Page #342 background image
F
SCPI Conformance Information
F-1
Introduction
The Model 7001 complies with SCPI version 1991.0.
Table F-1
Syntax of SCPI confirmed commands implemented by Model 7001
Command Description
:DISPlay
[ :WINDow[1]]
:TEXT
:DATA <a>
:STATe <b>
:STATe?
:WINDow2
:TEXT
:DATA <a>
:STATe <b>
:STATe?
Path to locate message to top portion of display
Path to control user text messages:
DeÞne ASCII message ÒaÓ using up to 20 characters.
Enable (1 or ON) or disable (0 or OFF) message.
Request message status (enabled or disabled).
Path to locate message to bottom portion of display:
Path to control user text messages:
DeÞne ASCII message ÒaÓ using up to 32 characters.
Enable (1 or ON) or disable (0 or OFF) message.
Request message status (enabled or disabled).
[:ROUTe]
:CLOSe <list>
:STATe?
:OPEN <list>|ALL
:SCAN <list>
:POINts?
Path and command to close speciÞed channels:
Request channels that are closed.
Open speciÞed (or all) channels.
Path and command to specify channel list:
Request number of channels in the scan list.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
BrandKeithley
Model7001
CategorySwitch
LanguageEnglish

Related product manuals