IEEE-488 Reference
5-87
[:ROUTe] subsystem
:STIMe <n>
[:ROUTe]:CONFigure:SLOT[1]:STIMe <n> Set delay for Slot 1
[:ROUTe]:CONFigure:SLOT2:STIMe <n> Set delay for Slot 2
<n> = 0 to 99999.999 Settling time in seconds
= DEFault 0 sec
= MINimum 0 sec
= MAXimum 99999.999 sec
:conf:slot:stim <n>
:conf:slot2:stim <n>
Power-up 0 sec
*RST 0 sec
:SYSTem:PRESet 0 sec
:STIMe? Query programmed delay for specified slot
:STIMe? DEFault Query *RST default delay
:STIMe? MINimum Query minimum delay
:STIME? MAXimum Query maximum delay
Short-form formats: :conf:slot:stim? :conf:slot2:stim?
:conf:slot:stim? def :conf:slot2:stim? def
:conf:slot:stim? min :conf:slot2:stim? min
:conf:slot:stim? max :conf:slot2:stim? max
Response message: 0 to 99999.999
An internally set relay settling time is provided to allow for switch bounce when it closes. At
power-up, a settling time is automatically set based on the acquired identification information
from the card. The :STIMe command allows you to increase the delay for each slot.
The delay value can be entered using the exponent form. For example, instead of sending a pa-
rameter value of 2000 (seconds), you can instead send 2E3. To cancel this delay, send this com-
mand using a parameter value of 0.
For complete information on this delay, see DELAY in Section 4.
10 OUTPUT 707; “:conf:slot:stim 10; stim?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Sets a 10 second delay for Slot 1 and then queries the delay period.
Line 20 Addresses the 7001 to talk.
Line 30 Displays the delay period (10).