IEEE-488 Reference
5-43
5.10.4
*ESR?
event status register query
To read the contents of the Standard Event Status Register and clear it.
*ESR?
Power-up Clears register
*CLS Clears register
*RST No effect
:STATus:PRESet No effect
DCL, SDC No effect
This command is used to acquire the value (in decimal) of the Standard Event Status Register
(see Figure 5-14). The binary equivalent of the returned decimal value determines which bits in
the register are set. When the *ESR? command is sent, the decimal value is placed in the Output
Queue and the Standard Event Status Register is cleared. When the Model 7001 is addressed to
talk, the value in the Output Queue is sent to the computer.
A set bit in this register indicates that a particular event has occurred. For example, for an ac-
quired decimal value of 48, the binary equivalent is 0011 0000. For this binary value, bits B4
and B5 of the Standard Event Status Register are set. These set bits indicate that a device-de-
pendent error and command error have occurred.
The bits of the Standard Event Status Register are described as follows:
Bit B0, Operation Complete (OPC)
Set bit indicates that all pending selected device oper-
ations are completed and the Model 7001 is ready to accept new commands. This bit will only
set in response to the *OPC? query command (see paragraph 5.10.7).
Bit B1
Not Used
Bit B2, Query Error (QYE)
Set bit indicates that you attempted to read data from an empty
Output Queue.
Bit B3, Device-dependent Error (DDE)
Set bit indicates that an instrument operation did
not execute properly due to some internal condition.
Bit B4, Execution Error (EXE)
Set bit indicates that the Model 7001 detected an error while
trying to execute a command.
Bit B5, Command Error (CME)
Set bit indicates that a command error has occurred. Com-
mand errors include:
1. IEEE-488.2 syntax error: Model 7001 received a message that does not follow the defined
syntax of the IEEE-488.2 standard.
Purpose
Format
Defaults
Description
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
BrandKeithley
Model7001
CategorySwitch
LanguageEnglish

Related product manuals