Message exchange protocol
The message exchange protocol can be summarized by the two following rules:
Rule 1. You must always tell the Model 7001 what to send to the computer.
The following two steps must always be performed, in this order, to send information
from the instrument to the computer:
1. Send the appropriate query command(s) in a program message.
2. Address the Model 7001 to talk.
Rule 2. The complete response message (including the RMT) must be received by
the computer before another program message can be sent to the Model
7001.