IEEE-488 Reference
5-38
Table 5-4
IEEE-488.2 common commands and queries
Mnemonic Name Description Para.
*CLS Clear status Clears all event registers, and Error Queue. 5.10.1
*ESE <NRf> Event status
enable command
Sets the contents of the Standard Event Status Enable Reg-
ister.
5.10.2
*ESE? Event status enable query Request the programmed value of the Standard Event Status
Enable Register.
5.10.3
*ESR? Event status
register query
Request the programmed value of the Standard Event Status
Register and clears it.
5.10.4
*IDN? Identification query Request the manufacturer, model number, serial number,
and firmware revision levels of the unit.
5.10.5
*OPC Operation complete
command
Sets the Operation Complete bit in the Standard Event Sta-
tus Register after all pending commands have been exe-
cuted.
5.10.6
*OPC? Operation complete query Places an ASCII “1” into the output queue when all pending
selected device operations have been completed.
5.10.7
*OPT? Option identification query Request the assigned model number for all slots. 5.10.8
*RCL <NRf> Recall command Returns the Model 7001 to the setup configuration stored in
the designated memory location.
5.10.9
*RST Reset command Returns the Model 7001 to the *RST default conditions. 5.10.10
*SAV <NRf> Save command Saves the current setup configuration to the designated
memory location.
5.10.11
*SRE <NRf> Service request
enable command
Sets the contents of the Service Request Enable Register. 5.10.12
*SRE? Service request
enable query
The Model 7001 returns the value of the Service Request
Enable Register.
5.10.13
*STB? Read status byte query Returns the value of the Status Byte Register. 5.10.14
*TRG Trigger command This command issues a bus trigger which has the same
effect as group execute trigger (GET) command.
5.10.15
*TST? Self-test query When this query is sent, the Model 7001 will perform a
checksum test on ROM and return the results.
5.10.16
*WAI Wait-to-continue
command
Wait until all previous commands are executed. 5.10.17
Note: At least one space is required between the command and the parameter.