IEEE-488 Reference
5-63
Line 70 Wait until scan is finished.
Line 80 Close channels 11 through 20.
When the above program is run, channels 1 through 10 will be scanned and then, at the end
of the scan, channels 11 through 20 will close. Without the *WAI command (Line 70),
channels 11 through 20 will close at the same time that the scan starts.