EasyManuals Logo

Keithley 7001 User Manual

Keithley 7001
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #188 background imageLoading...
Page #188 background image
IEEE-488 Reference
5-69
:STATe <b>
:DISPlay[:WINDow[1]]:TEXT:STATe <b> Control message; top display
:DISPlay:WINDow2:TEXT:STATe <b> Control message; bottom display
<b> = 0 or OFF Disable text message for specified display
= 1 or ON Enable text message for specified display
:disp:text:stat <b>
:disp:wind2:text:stat <b>
Power-up OFF
*RST No effect
:SYSTem:PRESet No effect
:STATe? Query state of message mode for specified display
Short-form formats: :disp:text:stat?
:disp:wind2:text:stat?
Response message: 1 (on) or 0 (off)
These commands enable and disable the text message modes. When enabled, a defined message
(see previous command) will be shown on the top or bottom portion of the display. When dis-
abled, the message will be removed from the display, and the Model 7001 will resume normal
operation.
A user defined text message remains displayed only as long as the instrument is in remote. Tak-
ing the instrument out of remote (by pressing the LOCAL key or sending LOCAL 707), cancels
the message and disables the text message mode.
10 OUTPUT 707; “:disp:text:stat on; stat?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Displays a defined user message (see previous command) on the top display, and then
queries the state (on or off) of the display.
Line 20 Addresses the 7001 to talk.
Line 30 Displays (on the CRT) the state of the message mode (1; on).
Parameters
Short-form
formats
Defaults
Query
command
Description
Programming
example
:DISPlay subsystem
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
BrandKeithley
Model7001
CategorySwitch
LanguageEnglish

Related product manuals