IEEE-488 Reference
5-71
Line 10 Disables the LED Graticule and then queries the state of the LED Graticule.
Line 20 Addresses the 7001 to talk.
Line 30 Displays (on the CRT) the state of the LED Graticule (0; off).
:SMESsage <b>
:DISPlay:SMESsage <b> Control status messages
<b> = 0 or OFF Disable status messages
= 1 or ON Enable status messages
:disp:smes <b>
Power-up Saved power-on setup
*RST OFF
:SYSTem:PRESet OFF
:SMESsage? Query state of status message mode
Short-form formats: :disp:smes?
Response message: 1 (on) or 0 (off)
This command is used to enable/disable the status message mode for the Model 7001. The status
message mode is a diagnostic tool that provides real-time messages that relate to the current op-
erating state of the instrument. When the status message mode is disabled, only the error events
(EE) in Table 4-1 will be reported to the display.
10 OUTPUT 707; “:disp:smes on; smes?”
20 ENTER 707; A$
30 PRINT A$
40 END
Line 10 Enables the status message mode, and then queries the state of the message mode.
Line 20 Addresses the 7001 to talk.
Line 30 Displays the state of the status message mode (1; on).
Parameters
Short-form
formats
Defaults
Query
command
Description
Programming
example
:DISPlay subsystem
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Other manuals for Keithley 7001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 7001 and is the answer not in the manual?

Keithley 7001 Specifications

General IconGeneral
BrandKeithley
Model7001
CategorySwitch
LanguageEnglish

Related product manuals