MLS-BA-defsi-0713
92
Ö Temperature calibration for the second point will
then begin.
After 14:59 minutes, temperature calibration for the
second point is complete.
The temperature value is adjusted and saved
automatically.
The balance returns to weighing mode
automatically.
L
L
Note:
The colour of the measuring sensor must be identical to that of the sensor point (e.g.
black sensor point - black measuring sensor)!
Sensor
point