EasyManua.ls Logo

Keysight Technologies 34420A

Keysight Technologies 34420A
189 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Calibration Procedures
100 Keysight 34420A Service Guide
3 Connect a multifunction calibrator to the meter's input terminals using the
provided cable (Keysight PN 34102A). Ensure the connector coupling nut is
properly tightened.
4 Allow 5 minutes for the input connector to temperature stabilize.
5 Select each function and range in the order shown in Table 3-4. Apply the
input signal shown.
6 Turn on the menu and select G:CALIBRATE menu.
7 Move to the commands level and select 2:CALIBRATE.
8 Move down to the parameters level and set the value in the display to the
calibrator input value.
9 Execute the command to begin the gain calibration.
10 Repeat step 1 through step 9 for each function and gain listed in the table.
11 Perform the “Gain Verification Procedure” on page 84 to check the results.
Table 3-4 Gain calibration procedure
Step#
Calibrator Output
[a]
[a] Gain calibration can be performed with an input that is within 0.9 to 1.1 of range.
Function/Channel Range
1
100.00000 mV
[b]
[b] The constants calculated for these ranges are used to derive the gain calibration constants for
the 1 mV and 10 mV ranges.
DCV / Channel 1 100mV
2
1.0000000 V
[b]
DCV / Channel 1 1 V
3
10.000000 V
[b]
DCV / Channel 1 10 V
4 100.00000 V DCV / Channel 1 100V
5 1.000,0000 4-wire Ohms 1
6 10.000000 4-wire Ohms 10
7 100.00000 4-wire Ohms 100
8 1,000.0000 4-wire Ohms 1 k
9 10,000.000 4-wire Ohms 10 k
10 100,000.00 4-wire Ohms 100 k
11 1,000,000.0 4-wire Ohms 1 M

Table of Contents

Related product manuals