EasyManua.ls Logo

Keysight Technologies 34420A

Keysight Technologies 34420A
189 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Calibration Procedures
86 Keysight 34420A Service Guide
1 mV and 10 mV Gain Verification (Optional)
This optional procedure checks the “full scale” gain for the 1 mV and 10 mV
ranges. Because of the low voltage levels used to verify these ranges and the
inherent inaccuracies of multifunction calibrators at these levels, a voltage divider
is used to check these ranges.
1 mV and 10 mV gain verification procedure
1 Set the meter to the following configuration:
Channel 1 input
7 ½ digits (100 PLC)
Filters OFF
2 Connect the calibrator, resistive divider, and meter as shown.
3 Measure or calculate the divider ratio.
NOTE
These ranges can only be verified. The gain calibrations for these ranges
are derived from the calibrations on the 100mV, JV, and 10 V ranges.
Calibrator
Meter
Source +
Sense +
Source -
Sense -
1 k
1
Isothermal
environment
HI
LO

Table of Contents

Related product manuals