Calibration Procedures 3
Keysight 34420A Service Guide 83
15 Low Power Ohms 10 20 µ 20 µ 20 µ
16 Low Power Ohms 100 200 µ 200 µ 200 µ
17 Low Power Ohms 1 k 2 m 2 mn2 m
18 Low Power Ohms 10 k 40 m 40 m 40 m
19 Low Power Ohms 100 k 1.2 1.5 1.5
Q 20 Low Power Ohms 1 M 3 4 4
21 Voltage Limited Ohms 10 20 µ 20 µ 20 m
22 Voltage Limited Ohms 100 200 µ 200 µ 200 µ
Optional
[b]
2-wire Ohms All
Use 4-wire Ohms specifications
[b]
[a] This value includes 100 nV for NULL off, see the specifications in datasheet.
[b] For 2-wire Ohms with Null ON (within ± 1°C of NULL), add 2 m to the listed specification. If null is not used, add 0.2 to the listed
specification.
Table 3-1 Zero Offset Verification Procedure
± Error From Zero
Step# Function/Channel Range 24 hour 90day 1 year