3-16 Installation and Service Guide N5250-90001
System Installation System
Cable Connections
PNA Series Microwave Network Analyzer System
N5250A
Test IF
N5260-60004
A IF
N5260A
SMA
(48” length)
85105-60033
Ref IF
N5260-60004
R1 IF
N5260A
SMA
(48” length)
85105-60033
RF In
N5260-60004
RF Out
N5260A
3.5 mm
(48” length)
8121-1221
LO In
N5260-60004
LO Out
N5260A
3.5 mm
(48” length)
8121-1221
RF In
N5260-60001 (N5250C Standard)
Test Port 2
E8361C H11
1.85 mm
(30” length)
N5260-60009
RF In N5260-60012 (N5250C Options
017/018)
Source Out 2
E8361C H11
1.85 mm
(30” length)
8121-1233
Rcvr B
N5260-60001 or N5260-60012
Rcvr B
E8361C H11
1.85 mm
(30” length)
8121-1233
From Left Millimeter-Wave
Test Head Module:
To: Cable Type Part Number
LO In
N5260-60003
LO Out
3.5 mm
(48” length)
8121-1221
RF IF
N5260-60003
RF Out
3.5 mm
(48” length)
8121-1221
Ref IF
N5260-60003
R2 IF
SMA
(48” length)
85105-60033
Test IF
N5260-60003
B IF
SMA
(48” length)
85105-60033
Bias (+12 V)
N5260-60003
Bias
Multi pin
(48” length)
85105-60030
RF In
N5260-60002 (N5250C Standard)
Test Port 1
E8361C H11
1.85 mm
(30” length)
N5260-60009
RF In N5260-60013 (N5250C Options
017/018)
Source Out 1
E8361C H11
1.85 mm
(30” length)
8121-1233
Rcvr A
N5260-60002 or N5260-60013
Rcvr A
E8361C H11
1.85 mm
(30” length)
8121-1233
Table 3-5 N5250C Cable List and Connections (Continued)