6-4 Installation and Service Guide N5250-90001
Performance Tests and Checks
System Check
PNA Series Microwave Network Analyzer System
N5250A
S
ystem Check
1. Connect a 1.0 mm sho
rt to Port 1 and Port 2. An adapter may be used on Port 2.
2. Restart the PNA application and perform a factory preset by pressing the PRESET key.
3. On the PNA menu bar, select: System > Service > Utilities > Receiver Display.
4. Set the IF bandwidth to 1 kHz.
The traces on the display should be similar to the example traces shown in Figure 6-1. If there
are power holes o
r other unexpected characteristics, examine the system for loose or damaged
cables, dirty or damaged connectors, proper connector torque, etc.
Figure
6-1
Typical Receiver Display
When you have a display that represents the current system performance, capture the display in
a file and save the file in the directory D:\sysver results (create this directory if it does not
already exist). To capture the display in a file, select: File > Print. An example of a recommended
file name is “opcheck_Sep_15_2016.png”.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5250C PNA Series and is the answer not in the manual?

Keysight Technologies N5250C PNA Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5250C PNA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals