Installation and Service Guide N5250-90001 4-5
PNA Series Microwave Network Analyzer System
N5250A
System Specifications
Specifications
N5250C Option 017 and 018 Test Head Bias-Tees Specifications
a. Pulse input connectors are operational only with Option H08 (Pulsed
Measurement Capabil
ity) enabled.
b. Based on deviation from signal reduction equation: Signal Reduction
(dB) = 20log
10
(Duty_cycle) = 20log
10
(pulse_width/period). Measured at Pulse
Repetition Frequency (PFR) of 1 MHz.
c. For rear panel LO port frequency, divide by 8.
d. For rear panel RF port frequency, divide by 6.
Table 4-3 Test Head Bias Input Connector
GNDU:
Sub Mini-Triaxial Connector, Trompeter BJ152 Insulated Bulkhead
Jack (150 Series)
Sense:
Sub Mini-Triaxial Connector, Trompeter BJ152 Insulated Bulkhead
Jack (150 Series)
Force:
Maximum Vol
tage:
Damage Voltage:
Maximum Current:
Sub Mini-Triaxial Connector, Trompeter BJ152 Insulated Bulkhead
Jack (150 Series)
± 30 VDC (typical)
± 40 VDC
± 0.5 AMP