Installation and Service Guide N5250-90001 5-3
PNA Series Microwave Network Analyzer System
N5250A
Maintenance and Support
Principles of Connector Care
Principles of Connector Care
Proper connector care and connection techniques are critical for accurate and repeatable
measurements. Refer to Table 5-1 for tips on connector care.
Prior to making connections to your analyzer, carefully review the information about
inspecting, cleaning, and gaging connectors. Refer to the calibration kit documentation for
detailed connector care information.
For course numbers about additional connector care instruction, contact Keysight
Technologies. Refer to Contacting Keysight on page 5-4.
Table 5-1 Connector Care Quick Reference Guide
Handling and Storage
Do Do Not
Keep connectors clean
Extend sleeve or connector nut
Use plastic end-caps during storage
Touch mating-plane surfaces
Set connectors contact-end down
Store connectors or adapters loose
Visual Inspection
Do Do Not
Use a damaged connector - ever Inspect all connectors carefully
Look for metal particles, scratches, and dents
Connector Cleaning
Do Do Not
a. Cleaning connectors with alcohol shall only be
done with the instruments power cord removed,
and in a well-ventilated area. Allow all residual alcohol moisture to evaporate, and the fumes
to dissipate, prior to energizing the instrument.
Use any abrasives
Get liquid into plastic support beads
Try compressed air first
Use isopropyl alcohol
a
Clean connector threads
Gaging Connectors
Do Do Not
Use an out-of-specification connector Clean and zero the gage before use
Use the correct gage type
Use correct end of calibration block
Gage all connectors before first use
Making Connections
Do Do Not
Align connectors carefully
Make preliminary connection contact lightly
Turn only the connector nut
Use a torque wrench for final connection
Apply bending force to connection
Over tighten preliminary connection
Twist or screw any connection
Tighten past torque wrench “break” point

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5250C PNA Series and is the answer not in the manual?

Keysight Technologies N5250C PNA Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5250C PNA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals