6-6 Installation and Service Guide N5250-90001
PNA Series Microwave Network Analyzer System
N5250A
Performance Tests and Checks
System Performance Verification
Materials Required
The following mate
rials are required to run the tests:
N5250C system (including system cable
s)
85059A 1.0 mm precision calibration and verification kit
General Preparation
Prepare for perfo
rmance verification by completing the following procedure:
1. Measure the
environment temperature and humidity. The temperature must be between +20 C
and +26 °C. Additionally, the temperature cannot vary by more than ±1 C after calibration.
2. Power on the system components in the following order:
a. N5260A millimeter head controller
b. E8361C Option H11 PNA
3. Remember to allow at least 90 minutes for warm up and temperature stabilization of the components.
The temperature of the calibration and verification kit must be stable with the environment temperature.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5250C PNA Series and is the answer not in the manual?

Keysight Technologies N5250C PNA Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5250C PNA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals