[SOURce:]ARB:FUNCtion:TYPE CURRent|VOLTage
[SOURce:]ARB:FUNCtion:TYPE?
Specifies either a voltage or current Arb. Only one type of Arb may be output at a time. The selection
must match the priority mode.
Parameter Typical Return
CURRent|VOLTage, *RST VOLTage VOLT or CURR
Specifies a voltage Arb: ARB:FUNC:TYPE VOLT
[SOURce:]ARB:TERMinate:LAST 0|OFF|1|ON
[SOURce:]ARB:TERMinate:LAST?
Selects the output setting after the Arb ends. When ON (1), the output voltage or current remains at
the last Arb value. The last Arb voltage or current value becomes the IMMediate value when the ARB
completes. When OFF (0), and also when the Arb is aborted, the output returns to the settings that
were in effect before the Arb started.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Terminatewith the output atthe last Arb value:
ARB:TERM:LASTON
6 SCPI Programming Reference
213 Keysight N6900/N7900 Series Operating and Service Guide