118 Chapter 7
Technical Characteristics
Waveform length
8 MS per channel (16 MS with option 016)
Minimum waveform length: 128 samples
Waveform granularity: 8 samples
Segments
From 1 to 32,768 unique segments can be defined consisting of waveform
start and stop address, repetitions and marker enable flags.
Sequences
Up to 16,384 total unique waveform segments can be combined with
separate loop counts to form a sequence.
External triggers
Front Panel
Triggers 1/2/3/4
Number of inputs: 4 SMB female front-panel connectors
Trigger polarity: Negative/positive
Trigger impedance: 2k ohms
Maximum input level: 4.3V
Input sensitivity: 250 mV
Trigger threshold: -4.5 to + 4.5 V
Trigger timing resolution: Clock/8 (6.4 ns at full rate)
Trigger uncertainty: < 50 ps
Minimum trigger width: 12.8 ns at full clock rate
Trigger delay: resolution of 1 SYNC clock

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N8241A and is the answer not in the manual?

Keysight Technologies N8241A Specifications

General IconGeneral
BrandKeysight Technologies
ModelN8241A
CategoryPortable Generator
LanguageEnglish

Related product manuals