122 Chapter 7
Technical Characteristics
General Characteristics
Power
Line power:
100/120/220/240 V AC, 50/60 Hz, 100 Watts maximum
Environmental
Samples of this product have been type tested in accordance with the Agilent
Environmental Test Manual and verified to be robust against the
environmental stresses of Storage, Transportation, and End-use; those
stresses include but are not limited to temperature, humidity, shock
vibration, altitude, and power line conditions. Test methods are aligned with
IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.
• Operating temperature: 0 to +50 degrees C (0 to 47 degrees C at 15,000 ft
altitude)
• Storage temperature: –40 to +70 degrees C
Relative Humidity
• Type tested: 0% to 95% at 40 degrees C (non-condensing)
• Altitude: 0 to 3000m (10,000 ft) above sea level
Weight
• 5.66 kg (12.5 lb)
Security
• All user data stored in volatile memory