Chapter 7 123
Technical Characteristics
Dimensions
• Height: 89 mm (3.5 inch)
• Width: 213 mm (8.375 inch)
• Depth: 422 mm (16.625 inch)
Recommended calibration cycle
• 12 months
ISO compliance
This instrument is manufactured in an ISO-9001 registered facility in
concurrence with Agilent Technologies, Inc. commitment to quality.
• Electrostatic Discharge Immunity - IEC 61000-4-2: passes criterion C
Testing according to IEC 61000-4-2: showed that the exposure of the data
port to electrostatic discharge may interrupt operation of the N8241/2A.
Operation can be restored by reloading the desired waveforms and
restarting. This required operator intervention results in a criterion C
classification.