:MAXimum? Returns the maximum or minimum value.
:MINimum?
:VOLTage
[:DC]? [<start_index>, <points>] Returns the averaged measurement.
:ACDC? Returns the RMS measurement (AC + DC).
:HIGH? Returns the High level of a pulse waveform.
:LOW? Returns the Low level of a pulse waveform.
:MAXimum? Returns the maximum or minimum value.
:MINimum?
:AHOur? [IGNORE_OVLD] Returns the accumulated amp-hours.
:ARRAY
:CURRent
[:DC]? [<start_index>, <points>] Returns the instantaneous measurement.
:POWer
[:DC]? Returns the instantaneous measurement.
:VOLTage
[:DC]? [<start_index>, <points>] Returns the instantaneous measurement.
:ELOG <maxrecords> Returns the most recent external datalog records.
:WHOur? [IGNORE_OVLD] Returns the accumulated watt-hours.
FORMat
[:DATA] ASCII | REAL Specifies the format of the returned data.
:BORDer NORMal | SWAPped Specifies how binary data is transferred.
HCOPy
:SDUMp
:DATA? [BMP|GIF|PNG] Returns an image of the front panel display.
:DATA
:FORMat BMP|GIF|PNG Specifies the format for front panel images returned.
IEEE-488 Common Commands
*CLS Clear status command.
*ESE<value> Event status enable command and query.
*ESR? Event status event query.
*IDN? Identification Query.
*OPC Sets the OPC (operation complete) bit in the standard event register.
*OPC? Returns a 1 to the output buffer when all pending operations complete.
*OPT? Returns a string identifying any installed options.
*RCL<value> Recalls a saved instrument state.
*RST Resets the instrument to pre-defined values that are either typical or safe.
*SAV<value> Saves the instrument state to one of ten non-volatile memory locations.
*SRE <value> Service request enable command and query.
*STB? Status byte query.
*TRG Trigger command.
Keysight RP7900 Series Operating and Service Guide 25
1 Quick Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies RP7900 Series and is the answer not in the manual?

Keysight Technologies RP7900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelRP7900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals