Keysight RP7933A/RP7943A
RP7933A/RP7943A
Test Record
Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
0.017996
Vout - 0.002004
19.994
Vout - 0.006
_______
_______
_______
_______
0.022004
Vout + 0.002004
20.006
Vout + 0.006
CV Load Effect:
Both - 0.001 _______ + 0.001
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
0.030
0.003
Transient response @ 300 µs:
Both - 0.8 _______ + 0.8
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
- 0.090
Iout - 0.090
799.59
Iout - 0.41
_______
_______
_______
_______
0.090
Iout + 0.090
800.41
Iout + 0.41
CC Load Effect:
Both - 0.050 _______ + 0.050
Current Sink Tests
100% of current rating: Both - 799.59 _______ - 800.41
RP7933A/RP7943A UUT Settings RP7933A/RP7943A Load Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
100% of current rating (Isink):
Voltage priority: 0.02 V, 408 A
Voltage priority: 20 V, 408 A
Voltage priority: 10 V, 816 A
Voltage priority: 10 V, 816 A
Current priority: 10.2 V, 0 A
Current priority: 10.2 V, 800 A
Current priority: 20.4 V, 400 A
Current priority: 10.2 V, - 800 A
not used
not used
Current priority: 10.2 V, 800 A
Current priority: 10.2 V, 400 A to 800 A,slew=10 A/µs
Voltage priority: 10 V, 0 A
Voltage priority: 10 V, 816 A
Voltage priority: 20 V, 408 A
Voltage priority: 10 V, 816 A
6 Verification and Calibration
284 Keysight RP7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies RP7900 Series and is the answer not in the manual?

Keysight Technologies RP7900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelRP7900 Series
CategoryPower Supply
LanguageEnglish

Related product manuals