53210A Math, Graphing, and Data Logging 6
Keysight 53210A User’s Guide 161
Clearing/Resetting Statistics
The current set of reading statistics is clear by any of the following:
enabling statistical functions
- CALCulate[1]:AVERage[:STATe] ON
a new measurement cycle - INITiate:IMMediate, READ?, MEASure?
sending a new SCPI command or changing a current SCPI parameter
a reset or instrument preset
- *RST, SYSTem:PRESet
These steps also clear all readings in reading memory. To clear statistics without
clearing reading memory, the following command is used:
CALCulate[1]:AVERage:CLEar[:IMMediate]
Limit Checking
Limit checking enables counter measurements to be compared on a
reading-by-reading basis to a lower limit and upper limit.
Limits that are exceeded are reported in the counter’s Questionable Data register
(bits 11 and 12). Register definitions and are contained in Chapter 8.
When enabled, limit checking appears on the counter display as shown in
Figure 6-4.