3 53210A Measurements
86 Keysight 53210A User’s Guide
Setting the Measurement Time Out
 
The measurement time-out is the time allowed for each measurement to 
complete. If a measurement does not complete before the time out expires, 
9.91E37 (not a number) is returned and the display indication is: - - - - - - -. The 
sequence continues with the next reading in the sample count. 
Specifying a time out prevents the instrument from pausing indefinitely if for some 
reason a measurement is unable to complete.
The measurement time out is set with the command:
SYSTem:TIMeout {<time>|MINimum|MAXimum|INFinity|DEFault} 
SYSTem:TIMeout? [{MINimum|MAXimum|DEFault}] (query form)
– time is the specified time out in the range of 10 ms to 2,000s with 1 ms 
resolution. The time out setting can affect reading throughput when multiple 
readings are taken during the trigger cycle. Trigger delays, gate times, and 
gate delays (Chapter 5) should be considered when setting an optimum time 
out value.
The time out setting is stored in non-volatile memory and is not changed 
following a reset (
*RST) or instrument preset (SYSTem:PREset or Preset key).
When shipped from the factory the measurement time out is set to 1 second. 
Specifying a time of 
9.9E+37 or sending the SYSTem:SECurity:IMMediate 
command also disables the time out. When the time out is disabled, the 
instrument will wait indefinitely for the measurement to complete.