EasyManua.ls Logo

Keysight 53210A - Keysight 53210 A Status System

Keysight 53210A
247 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Status 8
Keysight 53210A User’s Guide 223
Keysight 53210A Status System
This chapter provides an overview of the 53210A Status System. Refer to the
STATus subsystem and the IEEE-488 commands within the Programmer’s
Reference for bit definitions and additional information. The reference is located
on the 53210A/53220A/53230A Product Reference CD (p/n 53220-13601).
Questionable Data Register Group
The counter’s Questionable Data Register Group monitors the following
conditions and events within the counter:
input voltage overload
frequency measurement hardware time out
calibration data
time base error
lower-limit exceeded
upper limit exceeded
reading memory overflow
Condition Register
The Condition Register within the Questionable Data Register Group monitors
current conditions that may impact the integrity of the counter measurements.
The Condition Register is read with the command:
STATus:QUEStionable:CONDition?
Reading the register does not clear the bit(s) in the register. The bit is cleared
when the condition is removed.
Event Register
The Event Register within the Questionable Data Register Group also monitors
conditions that may impact the integrity of the counter measurements.

Table of Contents