6 53210A Math, Graphing, and Data Logging
190 Keysight 53210A User’s Guide
Graphics Functions and Reading Memory
Figure 6-13 summarizes the effects on reading memory when statistics and
histograms are reset, and when trend charts are cleared.
Figure 6-13 Reading Memory and Resetting/Clearing Graphics.
reading memory
.
.
.
.
.
.
(record)
.
.
.
(record)
(record)
Reset Hi st ogr am:
- clears current histogram (record)
- does not clear reading memory, does not abort readings
– clears count and new histogram re-starts on next reading
after reset
(record)
Clear Readings (Trend Chart):
- clears the trend chart
- clears reading memory, does not abort readings
- clears count and re-starts on next reading after clear
Reset St a t s:
- clears current set of statistics (record)
- does not clear reading memory, does not abort readings
– clears count and new set of statistics re-start on next
reading after reset
(memory)