6 53210A Math, Graphing, and Data Logging
174 Keysight 53210A User’s Guide
Resetting the Histogram
The data from which the current histogram is created is cleared by any of the
following:
pressing the Reset Histogram softkey
enabling/disabling or changing any part of the histogram
number of bins, lower or upper range
a new measurement cycle - INITiate:IMMediate, READ?, MEASure
sending a new SCPI command or changing a current SCPI parameter
an instrument reset or present
- *RST, SYSTem:PRESet
These actions also clear all readings in reading memory and in the output buffer.
To clear histogram data without clearing the readings in reading memory, the
following command is used:
CALCulate2:TARNsform:HISTogram:CLEar[:IMMediate]