A 53210A Error Messages
238 Keysight 53210A User’s Guide
+714
Channel 2 calibration signal power level
error
The power level of the calibration signal was outside the expected
range of the specified level.
+715
Channel 2 calibration signal frequency
error
The frequency of the calibration signal was outside the expect
range of the specified value.
+716
Channel 2 calibration signal
is not CW
The Channel 2 calibration signal must be a continuous wave (CW)
and not a pulsed signal.
+717
Channel 2 calibration
timeout occurred
Calibration at the specified power level did not complete within
the ~ 1.2 seconds allowed. Perform a self-test to verify operation
of the Channel 2 board and re-check the parameters of the input
signal.
+720
Auto-calibration failed;
input signal detected
No signal should be present on the input channel during an
auto-cal.
+742
Calibration data lost:
corrections
The instrument’s calibration data was lost due to a firmware
update that may have made the current calibration data invalid, or
due to a hardware failure. Recalibrate the instrument.
+748
Calibration memory
write failure
An error occurred as calibration data was being written to the
counter’s internal flash memory. Run the self-test to verify counter
operation and repeat the calibration procedure.
+750 Calibration data not restored
Following repair, calibration data was not restored as prompted
from the front panel.
Self-Test Errors
+901
Self-Test failed:
auto-calibration failure
A failure occurred during the auto-calibration portion of the
self-test. Repeat the self-test. If the auto-calibration failure
remains there is an issue with the instrument hardware. Contact
Keysight.
+903
Self-Test failed:
real-time clock setting lost
Reset the instrument clock using the ‘Utility’ -> ‘System Setup’ ->
‘Date/Time’ keys, or using the SYSTem:TIME or SYSTEM:DATE
commands. Re-run the self-test.
Table A-1 53210A Error Message Descriptions.