EasyManuals Logo

Keysight E3634A User Manual

Keysight E3634A
205 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #118 background imageLoading...
Page #118 background image
4 Remote Interface Reference
118 Keysight E3633A and E3634A User’s Guide
VOLTage:PROTection? {MINimum|MAXimum};
This query returns the overvoltage protection trip level presently programmed.
VOLT:PROT? MAX
and
VOLT:PROT? MIN
return the maximum and minimum
programmable overvoltage trip levels.
VOLTage:PROTection:STATe {0|1|OFF|ON}
This command enables or disables the overvoltage protection function. An
overvoltage condition can be cleared with the
VOLT:PROT:CLE
command after
the condition that caused the OVP trip is removed. At
*RST
, this value is set to
“ON”.
VOLTage:PROTection:STATe?
This query returns the state of the overvoltage protection function. The
returned parameter is “0” (OFF) or “1” (ON).
VOLTage:PROTection:TRIPped?
This query returns a “1” if the overvoltage protection circuit is tripped and not
cleared or a “0” if not tripped.
VOLTage:PROTection:CLEar
This command causes the overvoltage protection circuit to be cleared. After
this command, the output voltage is restored to the state it was in before the
protection feature occurred and the OVP trip level remains unchanged to the
value presently programmed. Before sending this command, lower the output
voltage below the trip OVP point, or raise the OVP trip level above the output
setting.
Note that the overvoltage condition caused by an external source must be
removed first before proceeding this command.