EasyManuals Logo

Keysight E3634A User Manual

Keysight E3634A
205 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
3 Front-Panel Operation
98 Keysight E3633A and E3634A User’s Guide
Calibration Message
You can use the calibration message feature to record calibration information
about your power supply. For example, you can store such information as the last
calibration date, the next calibration due date, the power supply’s serial number,
or even the name and phone number of the person to contact for a new
calibration.
You can record and read information in the calibration message from the remote
interface only.
The power supply should be unsecured before sending a calibration
message.
The calibration message may contain up to 40 characters.
The calibration message is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
Remote interface operation:
The following command strong shows how to store a calibration message.
CAL:STR <quoted string> Store the cal message
“CAL:STR ‘CAL 12-05-98’”