Specifications 8
Keysight E3633A and E3634A User’s Guide 201
Stability, ±(% of output + offset)
Following a 30-minute warm-up, with the output in the ON state according to the
operating mode (CC with load or CV), and with a change in the output over 8
hours under constant load, line, and ambient temperature.
Output voltage overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V
if the output control is set to less than 1 V. If the output control is set to 1 V or
higher, there is no overshoot.
Programming language
SCPI (Standard Commands for Programmable Instruments)
State Storage Memory
Three (3) user-configurable stored states
Recommended calibration interval
1 year
Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation to the
(+) output to the (+) sense and the (-) output and the (-) sense terminals.
±240 Vdc when connecting insulated shorting conductors to the (+) output
to the (+) sense and the (-) output and the (-) sense terminals.
Voltage 0.02% + 1mV
Current 0.1% + 1mA