Front-Panel Operation 3
Keysight E3633A and E3634A User’s Guide 71
Disabling the Output
The output of the power supply can be disabled or enabled from the front panel.
When the power supply is in the “Off” state, the OFF annunciator turns on and the
output is disabled. The OFF annunciator turns off when the power supply returns
to the “On” state. When the output is disabled, the voltage value is 0 volts and the
current value is 0.02 amps.
The output state is stored in volatile memory; the output is always disabled when
power has been off or after a remote interface reset.
– Front-panel operation:
You can disable the output by pressing key. This key toggles between
output “Off” and “On” states.
– Remote interface operation:
While the output is disabled, the range of selection keys, the control knob, the
resolution selection keys, and adjust selection key are still working. If the display
is in the meter mode, you cannot see the changes of output voltage and current
settings on the display when turning the knob. To see or check the changes
when the outputs are disabled, the display should be in the limit mode.
OUTP {OFF|ON} Disable or enable the output