3 Front-Panel Operation
74 Keysight E3633A and E3634A User’s Guide
System-Related Operations
This section gives information on topics such as self-test, error conditions, and 
front-panel display control. This information is not directly related to setting up 
the power supply but is an important part of operating the power supply.
Self-test
A power-on self-test occurs automatically when you turn on the power supply. 
This test assures you that the power supply is operational. This test does not 
perform the extensive set of tests that are included as part of the complete 
self-test described below. If the power-on self-test fails, the ERROR annunciator 
turns on.
–A complete self-test performs a series of tests and takes approximately 2 
seconds to execute. If all tests pass, you can have a high confidence that 
the power supply is operational.
–If the complete self-test is successful, “PASS” is displayed on the front 
panel. If the self-test fails, “FAIL” is displayed and the ERROR annunciator 
turns on. See the Service Guide for instructions on returning the power 
supply to Keysight Technologies for service.
– Front-panel operation:
To perform the complete front panel self-test, hold down the   key as you 
turn on the power supply and hold down the key until you hear a long beep. 
The self-test will begin when you release the key following the beep.
– Remote interface operation:
Returns “0” if the complete self-test passes or “1” if it fails.
*TST?